Table 3±20: IntelliFrame Mainframe error and event codes

Code

Message

 

 

0

No error. The error queue is empty. Every error or event has been read or

 

was purposely cleared.

 

 

±100

Command error.

 

 

±101

Invalid character.

 

 

±102

Syntax error.

 

 

±103

Invalid separator.

 

 

±104

Data type error.

 

 

±108

Parameter not allowed. More parameters than expected were received for

 

the header.

 

 

±109

Missing parameter.

 

 

±110

Command header error.

 

 

±112

Program mnemonic too long.

 

 

±113

Unidentified header.

 

 

±120

Numeric data error.

 

 

±141

Invalid character data.

 

 

±144

Character data too long.

 

 

±160

Block data error.

 

 

±161

Invalid block data.

 

 

±200

Execution Error.

 

 

±203

Command protected.

 

 

±213

Init Ignored. The measurement initiation was ignored because another

 

measurement was already in progress.

 

 

±220

Parameter error. A program data element related error occurred.

 

 

±221

Settings conflict. A legal program data element was parsed but could not be

 

executed due to the current state of the mainframe.

 

 

±222

Data out of range. The value of the data was outside of the legal range.

 

 

±223

Too much data.

 

 

±224

Illegal parameter value. The parameter value must be one of the exact

 

values from the command list.

 

 

±315

Configuration memory Lost. Some instrument setting may be lost.

 

 

±350

Queue Overflow. An error occurred but there was no room in the queue to

 

record the error.

 

 

±380

Calibration failed.

 

 

±381

Execution queue overflow. Internal error.

 

 

±410

Query INTERRUPTED.

 

 