Examples

3.Send the STATus:QUEStionable:ENABle command to enable bit 4 in the Questionable Enable register. If any problems occur with any of the slots, the corresponding bit will be set.

STAT:QUES:ENAB 16

Example 3: Summarize Status Information

The following example shows how to return a summary of the status information using the STAT:SEVent command.

Send the STAT:SEVent? query and note the returned information. For example, assume that the query returns a decimal value of 1073741947 (0X4000007b). Decoding the information indicates that both fans failed and that the temperatures for slots 0 through 3 exceeded the specified limits.

Since bit 30 is set, the overflow register should also be checked (STAT:S2EV?).

Example 4: Monitor the Power Supply Currents

See STATus subsystem on page 3±31.

Example 5: VXIMonitor

See VXIMonitor subsystem on page 3±155.

3±184

VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual