STATus Subsystem
STATus:QUEStionable:TEMPerature?
Returns the contents of the Temperature Event register. A value of one in a bit indicates the corresponding condition bit has transitioned from false to true. Unused bits always return a value of zero. The register is cleared by the *CLS command.
There is no command associated with this query.
Syntax | STATus:QUEStionable:TEMPerature[:EVENt]? | ||
Parameters |
|
|
|
| Parameter | Query response | |
|
| Not applicable | <NR1> |
|
|
|
|
|
|
|
|
|
| Blower status word bits | Definition |
|
|
|
|
|
| D0 | Slot 0 Delta Temperature |
|
|
|
|
|
| D1 | Slot 1 Delta Temperature |
|
|
|
|
|
| D2 | Slot 2 Delta Temperature |
|
|
|
|
|
| D3 | Slot 3 Delta Temperature |
|
|
|
|
|
| D4 | Slot 4 Delta Temperature |
|
|
|
|
|
| D5 | Slot 5 Delta Temperature |
|
|
|
|
|
| D6 | Slot 6 Delta Temperature |
|
|
|
|
|
| D7 | Slot 7 Delta Temperature |
|
|
|
|
|
| D8 | Slot 8 Delta Temperature |
|
|
|
|
|
| D9 | Slot 9 Delta Temperature |
|
|
|
|
|
| D10 | Slot 10 Delta Temperature |
|
|
|
|
|
| D11 | Slot 11 Delta Temperature |
|
|
|
|
|
| D12 | Slot 12 Delta Temperature |
|
|
|
|
|
| D13 | Ambient Temperature |
|
|
|
|
|
| D14 | Not Used |
|
|
|
|
|
| D15 | Not Used |
Reset Value |
|
|
|
0 |
| ||
Errors and Events |
| None |
|
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual | 3±83 |