STATus Subsystem
STATus:QUEStionable:VOLTage?
Returns the contents of the Voltage Event register. A value of one in a bit indicates the corresponding power supply failure event has been asserted. Unused bits always return a value of zero. The register is cleared by the *CLS command.
There is no command associated with this query.
Syntax | STATus:QUEStionable:VOLTage[:EVENt]? | ||
Parameters |
|
|
|
| Parameter | Query response | |
|
| Not applicable | <NR1> |
|
|
|
|
|
|
|
|
|
| Voltage status word bits | Definition |
|
|
|
|
|
| D0 | +24 V |
|
|
|
|
|
| D1 | +12 V |
|
|
|
|
|
| D2 | +5 V |
|
|
|
|
|
| D3 | +5 V Standby |
|
|
|
|
|
| D4 | +5 V External |
|
|
|
|
|
| D5 | ±2 V |
|
|
|
|
|
| D6 | ±5.2 V |
|
|
|
|
|
| D7 | ±12 V |
|
|
|
|
|
| D8 | ±24 V |
|
|
|
|
|
| D9 | Not Used |
|
|
|
|
|
| D10 | Not Used |
|
|
|
|
|
| D11 | Not Used |
|
|
|
|
|
| D12 | Not Used |
|
|
|
|
|
| D13 | Not Used |
|
|
|
|
|
| D14 | Not Used |
|
|
|
|
|
| D15 | Not Used |
Reset Value |
|
|
|
0 |
| ||
Errors and Events |
| None |
|
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual | 3±91 |