VXIMonitor Subsystem

VXIMonitor:TINTerval

VXIMonitor:TINTerval?

Sets or queries the time interval the VXI Monitor monitors the VXI bus. The VXI monitor monitors the bus for the specified number of seconds and reports any detected VXI Monitor event.

The factory default setting is 10 seconds.

Syntax

VXIMonitor:TINTerval <time>

 

VXIMonitor:TINTerval?

Parameters

 

 

 

<time>

 

Query response

 

<NRf>

 

<NR1>

 

1 N 2147483647

 

 

MIN

 

1

 

MAX

 

2147483647

Reset Value

 

 

Not Applicable

Errors and Events

±222, ªData Out of Rangeº

 

You attempted to set the interval to an illegal value.

 

±221, ªSettings Conflict.º

 

You attempted to set the interval while the VXI Monitor was executing.

Dependencies

None

 

 

Examples

Command: VXIM:TINT 100

 

Query:

VXIM:TINT?

 

Response:

100

 

Related Commands

VXIMonitor:REPeat

 

VXIMonitor:STATe