VXIMonitor Subsystem
VXIMonitor:TINTerval
VXIMonitor:TINTerval?
Sets or queries the time interval the VXI Monitor monitors the VXI bus. The VXI monitor monitors the bus for the specified number of seconds and reports any detected VXI Monitor event.
The factory default setting is 10 seconds.
Syntax | VXIMonitor:TINTerval <time> | ||
| VXIMonitor:TINTerval? | ||
Parameters |
|
|
|
<time> |
| Query response | |
| <NRf> |
| <NR1> |
| 1 ≤ N ≤ 2147483647 |
| |
| MIN |
| 1 |
| MAX |
| 2147483647 |
Reset Value |
|
| |
Not Applicable | |||
Errors and Events | ±222, ªData Out of Rangeº | ||
| You attempted to set the interval to an illegal value. | ||
| ±221, ªSettings Conflict.º | ||
| You attempted to set the interval while the VXI Monitor was executing. | ||
Dependencies | None |
|
|
Examples | Command: VXIM:TINT 100 | ||
| Query: | VXIM:TINT? | |
| Response: | 100 |
|
Related Commands | VXIMonitor:REPeat | ||
| VXIMonitor:STATe |
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual | 3±159 |