SYSTem Subsystem

SYSTem:COMMunicate:SERial:ERESponse

SYSTem:COMMunicate:SERial:ERESponse?

Sets or queries if error response messages automatically return to the RS-232 port. When the mode is ON, all error and event response messages will output directly to the RS-232 port; the error and event response messages will not be placed in the Error/Event queue. When the mode is OFF, all error and event messages will be reported to the Error/Event queue and will not be output to the RS-232 port directly.

If the mode is ON, all asynchronous error messages are placed in the Error/Event queue. The event registers will not be updated by any event message.

Asynchronous event messages include ªSelf Test Failedº and ªConfiguration

Memory Lost.º

Syntax

SYSTem:C￿MMunicate:SERial:ERESponse <Boolean>

 

SYSTem:C￿MMunicate:SERial:ERESponse?

Parameters

 

 

 

 

 

<Boolean>

 

Query response

 

 

<NRf>

 

<NR1>

 

 

N 0

 

1

 

0

 

0

 

 

ON

 

1

 

 

OFF

 

0

Reset Value

 

 

 

 

Not Applicable

Errors and Events

 

±224, ªIllegal Parameter Valueº

 

 

You attempted to set use an illegal parameter value.

Dependencies

 

None

 

 

Examples

 

Command: SYST:C￿MM:SER:ERES ￿FF

 

Query:

SYST:C￿MM:SER:ERES?

 

 

Response:

0

 

Related Commands

All SYSTem:COMMunicate:SERial commands

3±116

VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual