STATus Subsystem
STATus:OPERation:VXIMonitor?
Queries the contents of the VXI Monitor Event Register. A value of one in a bit indicates the corresponding condition bit changed from false to true. Unused bits always return a value of zero. The register is cleared by the *CLS command.
The VXI Monitor must be running (See VXIMonitor Subsystem on page 3±155) for these events to be recorded.
There is no command associated with this query.
Syntax | STATus:OPERation:VXIMonitor[:EVENt]? | ||
Parameters |
|
|
|
| Parameter | Query response | |
|
| Not applicable | <NR1> |
|
|
|
|
|
|
|
|
|
| VXI Monitor event word bits | Definition |
|
|
|
|
|
| D0 | BERR Occurred |
|
|
|
|
|
| D1 | SYSFAIL Occurred |
|
|
|
|
|
| D2 | ACFAIL Occurred |
|
|
|
|
|
| D3 | AS Missing |
|
|
|
|
|
| D4 | Not Used |
|
|
|
|
|
| D5 | Not Used |
|
|
|
|
|
| D6 | Not Used |
|
|
|
|
|
| D7 | Not Used |
|
|
|
|
|
| D8 | Not Used |
|
|
|
|
|
| D9 | Not Used |
|
|
|
|
|
| D10 | Not Used |
|
|
|
|
|
| D11 | Not Used |
|
|
|
|
|
| D12 | Not Used |
|
|
|
|
|
| D13 | Not Used |
|
|
|
|
|
| D14 | Not Used |
|
|
|
|
|
| D15 | Not Used |
Reset Value |
|
|
|
0 |
| ||
Errors and Events |
| None |
|
3±42 | VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |