Examples

This section describes a few brief examples using some of the commands described in this manual. These examples are intended to show how the commands from the Command section work together to complete a task.

Example 1: Monitor Power Supply Voltages

This example shows how to monitor the mainframe power supply voltages

1.Send the STATus:QUEStionable:VOLTage:ENABle command to enable the bits in the Voltage Enable register. This command lets you monitor all voltage limits.

STAT:QUES:VOLT:ENAB 32767

2.Send the STATus:QUEStionable:ENABle command to enable the bit 0 in the Questionable Enable register. If any problems occur with any of the power supplies, bit 0 of the register will be set.

STAT:QUES:ENAB 1

Example 2: Monitor the Delta Temperature

This example shows how to monitor the delta temperature in slots 0, 1, 2, and 3. The maximum allowable delta temperature for slots 0 and 1 is 10 °C; the maximum allowable delta temperature for slots 2 and 3 is 5 °C.

1.Send the STATus:QUEStionable:TEMPerature:LIMit command to set the temperature limits for the four slots (one command per slot):

STAT:QUES:TEMP:LIM DELTA0, 10 STAT:QUES:TEMP:LIM DELTA1, 10 STAT:QUES:TEMP:LIM DELTA2, 5 STAT:QUES:TEMP:LIM DELTA3, 5

2.Send the STATus:QUEStionable:TEMPerature:ENABle command to enable the bits in the Temperature Enable register. This command allows you to measure temperatures for the individual slots.

STAT:QUES:TEMP:ENAB #H0F

VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual

3±183