IEEE 488.2 Common Commands
*TST?
This query initiates an internal self test and returns a failure code. If more than one failure occurred, only the the unique numeric identifier of the first self test that failed is reported. A value of zero is returned if there were no failures.
Each internal self test has a unique identifier in the range of 1000 to 1999.
*TST? and TEST? are identical queries. There is no command associated with this query. You can obtain additional test results with the TEST:RESult:VER- Bose? command.
Syntax | *TST |
|
|
Parameters |
|
| |
<failure code> | Query response | ||
| Not Applicable | <NR1> | |
|
|
| 0 1000 ≤ N ≤ 1999 |
Reset Value |
|
| |
Not Applicable |
| ||
Errors and Events | None |
|
|
Dependencies | None |
|
|
Examples | Query: | *TST |
|
| Response: | 0 |
|
Related Commands | TEST |
|
|
TEST:RESults:VERBose?
3±170 | VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |