STATus Subsystem
STATus:S2EVent?
Overflow register for the STATus:SEVent summary event register. Returns the summary information on the power supply current events, total power (wattage) events, TTLTrigger events and IRQ events.
A one in a bit indicates the corresponding event has been asserted. Unused bits always return a value of zero. The query return a value of zero. The query returns a value of zero upon power on or when you send the *CLS command. Reading the events by this command will not clear the events.
There is no command associated with this query.
Syntax STTusS2EVent
Parameter | Query response |
|
|
Not Applicable | <NR1> |
|
|
Event status word bits | Definition |
|
|
D0 | +24 V Current |
|
|
D1 | +12 V Current |
|
|
D2 | +5 V Current |
|
|
D3 | Not Used |
|
|
D4 | Not Used |
|
|
D5 | ±2 V Current |
|
|
D6 | ±5PT2V Current |
|
|
D7 | ±12V Current |
|
|
D8 | ±24 V Current |
|
|
D9 | Total Power |
|
|
D10 | IRQ1 occurred |
|
|
D11 | IRQ2 occurred |
|
|
D12 | IRQ3 occurred |
|
|
D13 | IRQ4 occurred |
|
|
D14 | IRQ5 occurred |
|
|
D15 | IRQ6 occurred |
|
|
D16 | IRQ7 occurred |
|
|
D17 | TTL0trig occurred |
|
|
3±108 | VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |