Getting Started

Configuration

After installing the mainframe, you are ready to configure it for your applica- tions. If you have not already done so, install your VXIbus modules in the mainframe following the installation guidelines for your VXIbus module.

Module Selection and Design Guidelines

The IntelliFrame Mainframe is UL listed to UL 3111-1 and CAN/CSA C22.2 No. 1010.1 and is IEC 1010-1 compliant. You should use VXIbus modules with the mainframe that are listed or designed to limit electric shock and fire.

HInvestigate the possibility of fire hazards due to the availability of power from the mainframe exceeding 150 W per slot. Use similar considerations for VXIbus Modules that externally receive power greater than 150 W per slot.

HPrinted wiring boards should be flame rated UL 94V±1 or better.

HTo prevent possible electrical shock, limit the voltage to the mainframe interface or chassis to less than 60 VDC, 30 VRMS, or 42.4 Vpk.

HEnsure that measuring inputs provide a means of limiting unintended voltages that may reach the mainframe interface or chassis to less than 60 VDC, 30 VRMS, or 42.4 Vpk.

1±20

VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual