STATus Subsystem
STATus:QUEStionable:CURRent:CONDition?
Queries the contents of the Current Condition register. A value of one in a bit indicates the corresponding power supply is currently failing. Unused bits always return a value of zero. Definition of the bits in this register can be found in STATus:QUEStionable:CURRent? command.
Reading the register does not clear it.
There is no command associated with this query.
Syntax | STTusQUEStionableCURRentCONDition | |
Parameters |
|
|
Parameter | Query response | |
| Not Applicable | <NR1> |
Reset Value |
|
|
Not Applicable |
| |
Errors and Events | None |
|
Dependencies | None |
|
Examples | Query: STTQUESCURRCOND |
|
| Response:0 |
|
Related Commands | STATus:QUEStionable:CURRent? |
|
STATus:QUEStionable:CURRent:ENABle
3±66 | VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |