Adjustment Procedures

CALibration

CALibration?

This a service-related command. The command calibrates the exhaust temperature of each slot. While the mainframe executes the command, the command sets the CALibrating bit in the Operation Status Condition register.

The query performs the same function as the command but it also returns a pass or fail response. A return value of 0 means that the calibration has passed and a return value between 10 and 22 indicates a calibration failure in slots 0 through 12.

Syntax CALibration[:ALL]

CALibration[:ALL]?

Parameters

<Parameter>

Query response

 

Not Applicable

<NR1>

 

 

0

 

 

10 ± 22

 

 

 

Reset Value

Not Applicable

Errors and Events

±203, ªCommand Protectedº

 

You attempted to execute a command that is protected by a password. See the

 

SYSTem:PASSword command beginning on page 7±5 for more information.

Dependencies

None

 

Examples

Command: CAL

 

Query:

CAL?

 

Response:

0

Related Commands

None

 

7±4

VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual