Specifications

Table 4±4: Certifications and compliances

Characteristic

Description

 

 

 

 

 

EC Declaration of Conformity

Meets intent of Directive 89/336/EEC for Electromagnetic

 

Compatibility and Low Voltage Directive 73/23/ECC for Product

 

Safety. Compliance was demonstrated to the following

 

specifications as listed in the Official Journal of the European

 

Communities:

 

 

 

 

EMC Directive 89/336/EEC:

 

 

 

EN 55011

Class A Limits for Conductive Emissions.

 

EN60555-2

AC Power Line Harmonic Emissions.

 

EN 55011

Class A Limits for Radiated Emissions.

 

EN 50082-1 Immunity:

 

 

 

IEC 801-2

Electrostatic Discharge Immunity

 

IEC 801-3

RF Electromagnetic Field Immunity

 

IEC 801-4

Electrical Fast Transient/Burst Immunity

 

IEC 801-5

Power Line Surge Immunity

 

Low Voltage Directive 73/23/EEC:

 

 

 

EN61010-1

Safety requirements for electrical equipment

 

 

for measurement, control, and laboratory use

 

 

Australian Declaration of

Complies with Australian EMC Framework, under the Standard

Conformity (C-Tick Mark)

applicable product class 2064, governing radiated and

 

conducted emissions.

 

 

 

 

Immunity, Enclosure,

Up to 8 kV with no change to control settings or impairment of

Electrostatic Discharge

normal operation, and up to 15 kV with no loss of stored data or

(ESD)

damage that prevents recovery of normal operation by the user

 

(IEC 801-2).

 

 

 

 

 

Immunity, Enclosure, Radio

No mainframe failures when the mainframe is subjected to a

Frequency Electromagnetic

3 V/m electromagnetic field over the frequency range of 27 MHz

Field

to 500 MHz (IEC 801-3).

 

 

 

 

Immunity, Fast Transients,

No loss of stored data, change to control settings, degradation

Common Mode

of performance, or temporary loss of function will occur when the

 

mainframe is subjected to the following transients (IEC 801-4):

 

 

 

 

 

 

Port

Peak Voltage

Tr/Th

Rep Frequency

 

 

 

 

 

Signal & Control 0.5 kV

5/50 ns

5 kHz

 

DC Power

0.5 kV

5/50 ns

5 kHz

 

AC Power

1.0 kV

5/50 ns

5 kHz

 

 

 

 

 

VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual