Getting Started

Remote On/Standby Configuration

The IntelliFrame Mainframe allows you to power on and power off the mainframe from the rear panel (passive monitor connector) as well as the front panel On/Standby switch. Alternatively, you can disable the front panel On/Standby switch.

NOTE. If you decide to bypass the front panel On/Standby switch, you should configure the mainframe before installing the rackmount hardware. Refer to Appendix B: Special Configuration Options and then return to this section to continue the installation procedures.

The Enhanced Monitor allows you to power off the mainframe through VXI and RS-232 commands. If you apply external power to the Enhanced Monitor, you can you can also power on and off the mainframe using RS-232 commands.

NOTE. Disconnecting the power cord provides the mainframe's voltage discon- nect. Even when powered off by the On/Standby switch, the power supply circuits are connected to primary power.

Front Panel SYSReset IntelliFrame Mainframe provides you the ability to generate a backplane Activation SYSReset signal from the front panel. This feature allows you to reset the

complete VXI system without having to power down the mainframe. A SYSReset activation is accomplished by pushing the right side (nonlighted side) of the On/Standby power switch.

NOTE. If you decide to configure the front panel to SYSReset, you should configure the mainframe before installing the rackmount hardware. Refer to Appendix B: Special Configuration Options for activation of this feature and then return to this section to continue the installation procedures.

Rackmount Installation The procedures for installing the rackmount kits are similar for the Option 1R and 2R kits. You can install Option 1R kit to either recess the mainframe inside the rack or to protrude in front of the rack. However, you can only install the door for the Option 2R kit so that the mainframe is recessed inside the rack.

The steps for installing the rackmount kits consist of installing the rackmount hardware to the mainframe, and then installing the mainframe in the rack.

You will need the following tools to install the rackmount options:

HT-15 Torx driver

1±12