STATus Subsystem
STATus:SCONdition?
Returns the pass/fail condition of all the monitors. A value of one in a bit indicates the corresponding current condition has failed. Unused bits always return a value of zero. The definition of the bits can be found in STATus:SEVent? command on page 3±106.
Reading the register does not clear it.
There is no command associated with this query.
Summary information for current, power, triggers and IRQ reside in the overflow register. If bit 30 is set, the overflow register should be queried for more information. See the STATus:S2CONdition? query on page 3±105.
Syntax | STATus:SCONdition? |
| ||
Parameters |
|
|
|
|
| Parameter |
| Query response | |
|
| Not Applicable | <NR1> | |
Reset Value |
|
|
|
|
0 |
|
| ||
Errors and Events |
| None |
|
|
Dependencies |
| None |
|
|
Examples |
| Query: | STAT:SCON? |
|
|
| Response: | 256 |
|
Related Commands | STATus:SEVent? |
|
3±104 | VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |