STATus Subsystem

STATus:QUEStionable:TEMPerature:MAXimum?

Queries the maximum measurable temperatures in the mainframe since the last factory maintenance was performed; the returned value is in °C.

There is no command associated with this query.

Syntax

STATus:￿UESti￿￿able:TEMPerature:MAXimum? <L￿cati￿￿>

Parameters

 

 

 

 

 

<Location>

 

Query response

 

 

{AMBient OUT0 . . . OUT12 DELTa0 . . . DELTa12}

<NR1>

 

 

AMBient ± ambient input temperature

 

 

 

OUT0 . . . OUT12 ± output temperature of each slot

 

 

 

DELTa0 . . . DELTa12 ± output temperature minus

 

 

 

input temperature of each slot

 

Reset Value

 

 

 

 

0

 

 

Errors and Events

 

±224, ªIllegal Parameter Valueº

 

 

You attempted to use an illegal parameter value.

Dependencies

 

None

 

 

Examples

 

Query:

STAT:￿UES:TEMP:MAX? AMB

 

 

Response:

32

 

Related Commands

STATus:QUEStionable:TEMPerature:LEVel?

 

STATus:QUEStionable:TEMPerature:CMAXimum?

3±90

VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual