TEST Subsystem

TEST:RESults?

This query returns the results code for the last executed self test command. The unique numeric identifier of the first individual test that failed is reported. A value of zero is returned if there were no failures.

There is no associated command for this query.

Each internal self test has a unique numeric identifier in the range of 1000 to 1999 (see Table 3±19 on page 3±149).

Syntax

TEST￿RESults[￿CODE]￿

 

Parameters

 

 

 

Parameter

 

Query response

 

Not Applicable

<NR1>

 

 

 

0

 

 

 

1000 N 1999

Reset Value

 

 

Not Applicable

 

Errors and Events

None

 

 

Dependencies

None

 

 

Examples

Query:

TEST￿RES￿

 

 

Response:

0

 

Related Commands

TEST

 

 

TEST:RESults:VERBose?

VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual