STATus Subsystem
STATus:QUEStionable:VOLTage:PTR
STATus:QUEStionable:VOLTage:PTR?
Sets or queries the power supply positive transition filter.
You can only set or clear bits 3 and 4 of the transition filter. Setting a bit enables the corresponding power supply condition to propagate to the event system. Any failures will display on the front panel readout. If you disable the bits, any failure in the corresponding power supply will not be reported to the front panel readout or to the event system.
Bits 0 ± 2 and bits 5 ± 8 of the transition filter are always enabled. Bit 15 always returns a 0. The factory default setting is 32743. The +5 V Standby and +5 V External are not monitored.
Syntax | STATus:QUEStionable:VOLTage:PTR <Mask> | |||
| STATus:QUEStionable:VOLTage:PTR? |
| ||
Parameters |
|
|
|
|
| <Mask> |
| Query response | |
|
| {<NRf> <Nondecimal Numeric>} | <NR1> | |
|
| 0 ≤ N ≤ #HFFFF |
| |
|
|
|
| |
|
|
|
| |
|
| Voltage positive transition filter bits | Definition | |
|
|
|
|
|
|
| D3 |
| +5 V Standby |
|
|
|
|
|
|
| D4 |
| +5 V External |
Reset Value |
|
|
|
|
0 |
|
| ||
Errors and Events |
| ±222, ªData Out of Rangeº |
| |
|
| You attempted to set the transition filter to an illegal value. | ||
Dependencies |
| None |
|
|
Examples |
| Command: STAT:QUES:VOLT:PTR #H7FF7 | ||
|
| Query: | STAT:QUES:VOLT:PTR? |
|
|
| Response: | 32759 |
|
Related Commands STATus:QUEStionable:VOLTage:CONDition?
3±96 | VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual |