Adjustment Procedures
This chapter describes adjustment procedures for the IntelliFrame Mainframe. There are no hardware adjustments required for the IntelliFrame Mainframe. The information in this chapter describes the command set for service related tasks. Perform the calibration procedures when major modules have been replaced in the mainframe, such as the Enhanced Monitor, Backplane or Sense circuit boards.
The information in this section is designed for use by qualified service person- nel. Read the General Safety Summary and Service Safety Summary at the front of this manual and Service Strategy on page
NOTE. This procedure assumes that you have a PC running MS Windows with the Microsoft Terminal program.
The following steps describe the calibration procedure using the CALibration command:
1.Place the IntelliFrame Mainframe in a
2.Remove all modules from the IntelliFrame Mainframe.
3.Install the empty slot panel fillers upside down so the shutters are open and air can flow in all 13 slots.
4.Set the fan speed to the FULL position.
5.Power on the IntelliFrame Mainframe and let it run for five minutes.
6.While you are waiting for the mainframe to warm up, execute the Microsoft Terminal program.
The Microsoft Terminal program is located in the Accessories folder (for the default setup of Microsoft Windows).
7.Connect an
8.Select the Communications... command from the Settings menu of the Terminal program and select the appropriate COM port (for example, COM1 or COM2) from the dialog box.
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual | 7±1 |