IEEE 488.2 Common Commands

*ESE

*ESE?

Sets or queries the enable mask for the Standard Event Status Register. The

STATus:PRESet command does not change this enable mask.

 

The factory default setting is 0.

 

Syntax

*ESE <mask>

 

 

*ESE?

 

 

Parameters

 

 

 

Parameter

 

Query response

 

{<NRf> <Non-Decimal Numeric>}

<NR1>

 

0 N #HFF

 

 

 

 

 

 

 

 

Standard status byte definition

Query response

 

 

 

 

 

D0

 

Operation Complete

 

D1

 

Not Used

 

D2

 

Query Error

 

D3

 

Device Dependent Error

 

D4

 

Execution Error

 

D5

 

Command Error

 

D6

 

Not Used

 

D7

 

Power On

Reset Value

 

 

Not Applicable

 

Errors and Events

±222, ªData Out of Rangeº

 

 

You attempted to set the enable mask to an illegal value.

Dependencies

None

 

 

Examples

Command: *ESE #H40

 

 

Query:

*ESE?

 

 

Response:

64

 

Related Commands

*ESR?

 

 

3±162