13.Slide the mainframe back until it stops.

14.Push to release the button latches, located on the outside of each track, and continue to slide the mainframe all the way back into the rack cabinet.

15.Tighten the four screws that you left loose at the rear of the rack when you performed step 8, and pull the mainframe part way out of the rack. (Tighten the 10-32 screws using 28 inch-lbs of torque.)

16.Be sure that the screws that you left loose at the front of the rack are loose enough to allow the slide-out track assemblies to seek the normal positions.

17.Tighten the screws and push the mainframe all the way into the rack. If the tracks do not slide smoothly, readjust the level using the method described in steps 15 and 16. (Tighten the 10-32 screws using 28 inch-lbs of torque.)

18.Install the proper power cord at the rear of the mainframe.

NOTE. To ensure the mainframe is properly mounted in the rack do not tighten the screws while installing the chassis track to the rack front rails. After all the mounting screws have been installed shift the mainframe to the far right of the rack, then tighten all the mounting screws. Refer to Figure 1±10 for screw positions. Depending on the rack manufacturer it may be necessary to use an additional spacer plate, refer to the Replaceable Parts on page 9±15.

 

Left rack rail & chassis

Right rack rail &

 

slide assembly

chassis slide assembly

 

 

Rack front

Rack

 

rail

Spacer

Chassis

front rail

Chassis

plate

slide

 

assemblies

slide

 

 

assemblies

 

 

Shift the mainframe to the far right. Do not tighten the screws until all all screws are touching the left side of the slots.

Figure 1±10: R1 or R2 rackmounting screw positions (front view).

1±18