LICENSING AND ENABLING
All
Licensing the 8732 Diagnostics
TUNABLE EMPTY PIPE DETECTION
Tunable Empty Pipe Parameters
For licensing the advanced diagnostics, follow the steps below.
1.
2.Verify that you have 1.01.001 software or later
3.Determine the Device ID
4.Obtain a License Key from your local Rosemount Representative.
5.Enter License Key
6.Enable Advanced Diagnostics
The Tunable Empty Pipe detection provides a means of minimizing issues and false readings when the pipe is empty. This is most important in batching applications where the pipe may run empty with some regularity.
If the pipe is empty, this diagnostic will activate, set the flow rate to 0, and deliver a PlantWeb alert.
Turning Empty Pipe On/Of
The Empty Pipe diagnostic can be turned on or off as required by the application. If the advanced diagnostics suite 1 (D01 Option) was ordered, then the Empty Pipe diagnostic will be turned on. If D01 was not ordered, the default setting is off.
The Tunable Empty Pipe diagnostic has one
Empty Pipe Value
Reads the current Empty Pipe Value. This is a