Sensor Calibration Result

Displays the result of the sensor calibration verification test as pass or fail. For more details on this parameter see Appendix C: Diagnostics.

Coil Circuit Result

Displays the result of the coil circuit test as pass or fail. For more details on this parameter see Appendix C: Diagnostics.

Electrode Circuit Result

Displays the result of the electrode circuit test as pass or fail. For more details on this parameter see Appendix C: Diagnostics.

Sensor Signature

The sensor signature describes the sensor characteristics to the transmitter and is an integral part of the sensor meter verification test. From this menu you can view the current stored signature, have the transmitter take and store the sensor signature, and re-call the last saved good values for the sensor signature. For more details on this parameter see Appendix C: Diagnostics.

Signature Values

Review the current values stored for the sensor signature. For more details on this parameter see Appendix C: Diagnostics.

Coil Resistance

View the reference value for the coil resistance taken during the sensor signature process.

Coil Signature

View the reference value for the coil signature taken during the sensor signature process.

Electrode Resistance

View the reference value for the electrode resistance taken during the sensor signature process.

Re-Signature Meter

Have the transmitter measure and store the sensor signature values. These values will then be used as the baseline for the meter verification test. Use this when connecting to older Rosemount or competitors’ sensors or installing the magnetic flowmeter system for the first time. For more details on this parameter see Appendix C: Diagnostics.

Recall Last Saved Values

Recalls the last saved “good” values for the sensor signature.