Rosemount 8732
Reference Manual
Low Flow Cutoff
E-SERIES ADVANCED DIAGNOSTICS CAPABILITIES
Adjustable between 0.01 and 38.37 ft./s (0.003 and 11.7 m/s). Below selected value, output is driven to the zero flow rate signal level.
Overrange Capability
Signal output will remain linear until 110% of upper range value or 44 ft./s (13 m/s). The signal output will remain constant above these values. Out of range message displayed on LOI and the Field Communicator.
Damping
Adjustable between 0 and 256 seconds
Basic
Self test
Transmitter faults
Analog output test
Pulse output test
Tunable empty pipe
Reverse flow
Coil circuit fault
Electronics temperature
Process Diagnostics (D01)
Ground/wiring fault
High process noise
SMART Meter Verification (D02)
Smart Meter Verification
OUTPUT SIGNALS | 8732 LOI Lockout |
PROFIBUS PA FIELDBUS DIGITAL OUTPUT SPECIFICATIONS
All optical switches on the display can be locked locally from the display layout configuration screen by holding the upper right optical switch for 10 seconds. The display can be reactivated holding the same switch for 10 seconds.
Output Signal
Profile Version
3.01
Identification Number
Generic: 0x9740
Manufacturer Specific: 0x0C15