Rosemount 8732
Reference Manual
Basic Diagnostics
Advanced Diagnostics
The basic diagnostics menu contains all of the standard diagnostics and tests that are available in the 8732E transmitter.
Empty Pipe Limits
Empty Pipe allows you to view the current value and configure the diagnostic parameters. For more detail on this parameter see Appendix C: Diagnostics.
EP Value
Read the current Empty Pipe Value. This number is a unitless number and is calculated based on multiple installation and process variables. For more detail on this parameter see Appendix C: Diagnostics.
EP Trigger Level
Limits: 3 to 2000
Configure the threshold limit that the empty pipe value must exceed before the diagnostic alert activates. Default from the factory is set to 100. For more detail on this parameter see Appendix C: Diagnostics.
EP Counts
Limits: 5 to 50
Configure the number of consecutive times that the empty pipe value must exceed the empty pipe trigger level before the diagnostic alert activates. Counts are taken at 1.5 second intervals. Default from the factory is set to 5. For more detail on this parameter see Appendix C: Diagnostics.
Electronics Temp Value
Electronics Temperature allows you to view the current value for the electronics temperature.
The advanced diagnostics menu contains information on all of the additional diagnostics and tests that are available in the 8732 transmitter if one of the diagnostics suite packages was ordered.
Rosemount offers two advanced diagnostic suites. Functionality under this menu will depend on which of these suites are ordered.
Advanced diagnostics suite option one (D01 option) contains advanced diagnostics for High Process Noise detection and Grounding and Wiring fault detection.
Advanced diagnostics suite option two (D02 option) contains advanced diagnostics for the 8714i Meter Verification. This diagnostic is used to verify the accuracy and performance of the magnetic flow meter installation.
8714i Meter Verification
This diagnostic allows you to test and verify that the sensor, transmitter, or both are working within specifications. For more details on this diagnostic, see Appendix C: Diagnostics.