Reference Manual
Rosemount 8732
Figure 5-4. Incline or Decline Orientation
FLOW |
FLOW |
Figure 5-5. Horizontal Sensor Orientation
Horizontal installation should be restricted to low piping sections that are normally full. Orient the electrode plane to within 45 degrees of horizontal in horizontal installations. A deviation of more than 45 degrees of horizontal would place an electrode at or near the top of the sensor thereby making it more susceptible to insulation by air or entrapped gas at the top of the sensor.
FLOW |