Reference Manual

00809-0100-4665, Rev AA August 2010

Rosemount 8732

TRANSMITTER

TROUBLESHOOTING

Table 6-4. Advanced Troubleshooting–Rosemount 8732

Symptom

Potential Cause

 

Corrective Action

Does not appear to be within

Transmitter, control system, or other

Check all configuration variables for the transmitter, sensor,

rated accuracy

receiving device not configured

communicator, and/or control system

 

properly

Check these other transmitter settings:

 

 

 

 

•Sensor calibration number

 

 

•Units

 

 

•Line size

 

Electrode Coating

Use bulletnose electrodes;

 

 

Downsize sensor to increase flow rate above 3 ft/s;

 

 

Periodically clean sensor

 

Air in line

Move the sensor to another location in the process line to

 

 

ensure that it is full under all conditions.

 

Moisture problem

Perform the sensor Tests A, B, C, and D

 

 

(see Table 6-5 on page 6-8)

 

Improper wiring

If electrode shield and signal wires are switched, flow indication

 

 

will be about half of what is expected. Check wiring diagrams for

 

 

your application.

 

Flow rate is below 1 ft/s

See accuracy specification for specific transmitter and sensor

 

(specification issue)

 

 

 

Auto zero was not performed when the

Set the coil drive frequency to 37 Hz, verify the sensor is full,

 

coil drive frequency was changed from

verify there is no flow, and perform the auto zero function.

 

5 Hz to 37 Hz

 

 

 

Sensor failure–Shorted electrode

Perform the sensor Tests C and D

 

 

(see Table 6-5 on page 6-8)

 

Sensor failure–Shorted or open coil

Perform the sensor Tests A and B

 

 

(see Table 6-5 on page 6-8)

 

Transmitter failure

Verify transmitter operation with an 8714 Calibration Standard or

 

 

replace the electronic board

Noisy Process

Chemical additives upstream of

Complete the Noisy Process Basic procedure. Move injection

 

magnetic flowmeter

point downstream of magnetic flowmeter, or move magnetic

 

 

flowmeter.

 

Sludge flows–Mining/Coal/

Decrease flow rate below 10 ft/s

 

Sand/Slurries (other slurries with

 

 

 

hard particles)

 

 

 

Styrofoam or other insulating particles

Complete the Noisy Process Basic procedure;

 

in process

Consult factory

 

Electrode coating

Use replaceable electrodes in Rosemount 8705.

 

 

Use a smaller sensor to increase flow rate above 3 ft/s.

 

 

Periodically clean sensor.

 

Air in line

Move the sensor to another location in the process line to

 

 

ensure that it is full under all conditions.

 

Low conductivity fluids (below 10

Trim electrode and coil wires – see “Conduit Cables” on

 

microsiemens/cm)

 

page 2-6

 

 

• Keep flow rate below 3 FPS

 

 

Integral mount transmitter

 

 

Use 8712-0752-1,3 cable

 

 

• Use N0 approval sensor

 

Advanced Troubleshooting continued on next page

 

 

 

 

6-5

Page 69
Image 69
Emerson Rosemount 8732 manual Transmitter Troubleshooting, Symptom Potential Cause Corrective Action