Foxboro Sensor to 8732 Connect coil drive and electrode cables as shown in Figure E-15.

Transmitter

Figure E-15. Generic Wiring Diagram for Foxboro Sensors and Rosemount 8732

Table E-16. Foxboro Sensor Wiring Connections

FOXBORO

ROSEMOUNT 8732

TRANSMITTER

 

SENSOR

 

 

 

White

 

Electrodes

Black

 

 

Any Shield

 

 

Ground

 

Coils

L2

 

 

 

 

L1

 

 

Fuse

 

Rosemount 8732

 

Foxboro Sensors

1

 

 

 

L1

2

 

 

 

L2

 

 

 

 

 

 

Chassis Ground

 

 

 

 

 

17

 

 

Any Shield

18

 

 

Black

19

 

 

White

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This is a pulsed DC magnetic flowmeter.

Do not connect AC power to the sensor or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary.