Rosemount 8732

Reference Manual

00809-0100-4665, Rev AA August 2010

Index

Parameter

Description

26

DEVICE_MAN_ID

Mapped to Mfg_ID parameter in

 

 

Manufacturing Block

27

DEVICE_ID

Mapped to DEV_TYPE parameter in

 

 

Manufacturing Block

28

DEVICE_SER_NUM

Mapped to last 16 characters of

 

 

DEVICE_ID_STRING parameter in

 

 

Manufacturing Block

29

DIAGNOSIS

Detailed information of the device, bitwize

 

 

coded. More than one message is possible at

 

 

once. If MSB of byte 4 is set to 1 than more

 

 

diagnostic information is available in the

 

 

DIAGNOSIS_EXTENSION parameter.

30

DIAGNOSIS_EXTENSION

Additional manufacturer-specific information

 

 

of the device, bitwize coded. More than one

 

 

message is possible at once.

31

DIAGNOSIS_MASK

Supported std diagnostics

 

 

Warmstart

 

 

Coldstart

 

 

maint requd

 

 

ident number violation

 

 

function check

 

 

failed

 

 

maint demanded

 

 

extension available

32

DIAGNOSIS_MASK_EXTENSION

Definition of supported

 

 

DIAGNOSIS_EXTENSION information-bits.

33

RESERVED

 

34

WRITE_LOCKING

Software write protection.

 

 

0: acyclic write service of all parameter are

 

 

refused, except

 

 

WRITE_LOCKING and the TAB_ENTRY

 

 

parameter of the Linearization table, i.e.

 

 

access is denied.

 

 

2457: is the default value and means all

 

 

writable parameters of a device are writable.

35

FACTORY_RESET

1: (mandatory) is the command for resetting

 

 

device for default values. The setting of the

 

 

bus address is not affected.

 

 

2506: (optional) is the command for warmstart

 

 

of the device. All parametrisation remains

 

 

unchanged.

 

 

2712: (optional) The bus address to its default

 

 

address; other parametrisation remains

 

 

unchanged. The bus address is changed

 

 

immediately regardless if the

 

 

device is in cyclic data transfer state. The

 

 

reset is not suspended up to a subsequent

 

 

power cycle/warmstart.

36

DESCRIPTOR

User-definable text (a string) to describe the

 

 

device within the application.

37

DEVICE_MESSAGE

User-definable MESSAGE (a string) to

 

 

describe the device within the application or

 

 

in the plant.

38

DEVICE_INSTALL_DATE

Date of installation of the device.

39

LOCAL_OP_ENA

Local operation enable.

 

 

0: disabled (Local operation not allowed, i.e.

 

 

change of FB MODE from host

 

 

device only)

 

 

1: enabled (Local operation is allowed).

F-2

Page 154
Image 154
Emerson Rosemount 8732 manual Index Parameter Description, Diagnosis