PERFORMANCE SPECIFICATIONS
(System specifications are given using the frequency output and with the unit at reference conditions.)
Accuracy
Includes the combined effects of linearity, hysteresis, repeatability, and calibration uncertainty.
Rosemount E Series with 8705/8707 Sensor:
Standard system accuracy is ±0.25% of rate ±1.0 mm/sec from 0.04 to 6 ft./s (0.01 to 2 m/s); above 6 ft./s (2 m/s), the system has an accuracy of ±0.25% of rate ±1.5 mm/sec.
Optional high accuracy is ±0.15% of rate ±1.0 mm/sec from 0.04 to 13 ft./s (0.01 to 4 m/s); above 13 ft./s (4 m/s), the system has an accuracy of ±0.18% of rate.(1)
| 2.5 |
|
|
|
|
|
|
|
| 2.0 |
|
|
|
|
|
|
|
% of Rate | 1.5 |
|
|
|
|
|
|
|
1.0 |
|
| 0.25% |
| 0.15% |
|
| |
|
|
|
|
|
|
| ||
| 0.5 |
|
|
|
|
|
|
|
| 0 |
|
|
|
|
|
|
|
| 0 | 3 | 6 | 13 | 20 | 27 | 33 | 40 |
|
| (1) | (2) | (4) | (6) | (8) | (10) | (12) |
|
|
|
| Velocity in ft/s (m/s) |
|
|
Rosemount E-Series with 8711 Sensor:
Standard system accuracy is ±0.25% of rate ±2.0 mm/sec from 0.04 to 39 ft./s (0.01 to 12 m/s).
Optional high accuracy is ±0.15% of rate ±1.0 mm/sec from 0.04 to 13 ft./s (0.01 to 4 m/s); above 13 ft./s (4 m/s), the system has an accuracy of ±0.18% of rate.
| 2.5 |
|
|
|
|
|
|
|
| 2.0 |
|
|
|
|
|
|
|
% of Rate | 1.5 |
|
|
|
|
|
|
|
1.0 |
|
| 0.25% |
| 0.15% |
|
| |
|
|
|
|
|
|
| ||
| 0.5 |
|
|
|
|
|
|
|
| 0 |
|
|
|
|
|
|
|
| 0 | 3 | 6 | 13 | 20 | 27 | 33 | 40 |
|
| (1) | (2) | (4) | (6) | (8) | (10) | (12) |
|
|
|
| Velocity in ft/s (m/s) |
|
|
Rosemount E-Series with 8721 Sensor:
Standard system accuracy is ±0.5% of rate from 1 to 39 ft./s (0.3 to 12 m/s); between 0.04 and 1.0 ft./s (0.01 and 0.3 m/s), the system has an accuracy of ±0.005 ft./s (0.0015 m/s).
(1)For Sensor sizes greater than 12 in. (300 mm) the high accuracy is ±0.25% of rate from 3 to 39 ft./sec (1 to 12 m/sec).