Rosemount 8732
Reference Manual
SENSOR MOUNTING
Upstream/Downstream Piping
Figure 5-2. Upstream and
Downstream
Straight Pipe Diameters
Sensor Orientation
Figure 5-3. Vertical Sensor Orientation
Physical mounting of a sensor is similar to installing a typical section of pipe. Conventional tools, equipment, and accessories (bolts, gaskets, and grounding hardware) are required.
To ensure specification accuracy over widely varying process conditions, install the sensor a minimum of five straight pipe diameters upstream and two pipe diameters downstream from the electrode plane (see Figure 5-2).
5 Pipe Diameters | 2 Pipe Diameters |
FLOW |
|
The sensor should be installed in a position that ensures the sensor remains full during operation. Figures
Vertical installation allows upward process fluid flow and is generally preferred. Upward flow keeps the
Installations with reduced straight runs from 0 to five pipe diameters are possible. In reduced straight pipe run installations, performance will shift to as much as 0.5% of rate. Reported flow rates will still be highly repeatable.
FLOW |
FLOW |