Rosemount 8732
Reference Manual
Figure 2-3. Supply Current versus Input Voltage
Figure 2-3 shows the supply current for each corresponding supply voltage. For combinations not shown, you can calculate the maximum distance given the supply current, the voltage of the source, and the minimum start-up voltage of the transmitter, 12 V DC, using the following equation:
| MaximumResis tance = |
| |||||||||
|
|
|
|
|
|
|
|
| 1amp |
| |
|
|
|
|
|
|
|
|
|
|
|
|
| 0.9 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| 0.8 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
| 0.7 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
(Amps) | 0.6 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
0.5 |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
| |||
Current | 0.4 |
|
|
|
|
|
|
|
|
|
|
Supply | 0.3 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| 0.2 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
12 1824 30 36 42
Power Supply (Volts)
I = 10/V
I = Supply current requirement (Amps)
V = Power supply voltage (Volts)
Installation Category Overcurrent Protection
Connect Transmitter Power
The installation category for the Rosemount 8732 is (overvoltage) Category II.
The Rosemount 8732 Flowmeter Transmitter requires overcurrent protection of the supply lines. Maximum ratings of overcurrent devices are as follows:
Power System | Fuse Rating | Manufacturer |
250 V; 2 Amp, Quick Acting | Bussman AGCI or Equivalent | |
42 Vdc | 50 V, 3 Amp, Quick Acting | Bussman AGCI or Equivalent |
To connect power to the transmitter, complete the following steps.
1.Ensure that the power source and connecting cable meet the requirements outlined on page
2.Turn off the power source.
3.Open the power terminal cover.
4.Run the power cable through the conduit to the transmitter.
5.Connect the power cable leads as shown in Figure
a.Connect AC Neutral or DC- to terminal 9.
b.Connect AC Line or DC+ to terminal 10.
c.Connect AC Ground or DC Ground to the ground screw mounted inside the transmitter enclosure.