ENDRESS AND HAUSER Connect coil drive and electrode cables as shown in Figure E-5.

SENSORS

Endress and Hauser

Sensor to

Rosemount 8732

Transmitter

Figure E-5. Wiring Diagram for Endress and Hauser Sensors and Rosemount 8732

Table E-6. Endress and Hauser Sensor Wiring Connections

ROSEMOUNT 8732

TRANSMITTER

ENDRESS AND HAUSER

SENSORS

Electrodes

7

5

4

14

42

41

Coils

Rosemount 8732

 

Endress and Hauser Sensors

1

 

 

41

2

 

 

42

 

 

 

 

 

14

 

 

 

 

 

17

 

4

18

 

5

19

 

7

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This is a pulsed DC magnetic flowmeter.

Do not connect AC power to the sensor or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary.