Reference Manual

00809-0100-4665, Rev AA August 2010

Rosemount 8732

KENT VERIFLUX VTC SENSOR

Veriflux VTC Sensor to 8732 Transmitter

Connect coil drive and electrode cables as shown in Figure E-16.

Figure E-16. Wiring Diagram for Kent Veriflux VTC Sensor and Rosemount 8732

Table E-17. Kent Veriflux VTC Sensor Wiring Connections

KENT VERIFLUX VTC

ROSEMOUNT 8732

 

SENSOR

 

TRANSMITTER

Electrode Connections

 

3 4

2

 

 

SIG SIG 5 6

1SCR

 

 

1

OUT

 

 

2

 

 

 

5+ 6 SCR

1 2 -

Fuse

 

 

 

OUT

 

 

 

Coil Connections

 

Rosemount 8732

 

Kent Veriflux VTC Sensors

1

 

 

2

2

 

 

1

 

 

 

 

 

 

SCR OUT

 

 

 

 

 

17

 

 

SCR OUT

18

 

 

SIG1

19

 

 

SIG2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This is a pulsed DC magnetic flowmeter.

Do not connect AC power to the sensor or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary.