Rosemount 8732
Reference Manual
ADVANCED CONFIGURATION
DETAILED SETUP
In addition to the basic configuration options and the diagnostic information and controls, the 8732 has many advanced functions that can also be configured as required by the application.
The detailed setup function provides access to other parameters within the transmitter that can be configured such as coil drive frequency, output parameters, local display configuration, and other general information about the device.
Additional Parameters
The additional parameters menu provides a means to configure optional parameters within the 8732E transmitter.
Density Value
The density value is used to convert from a volumetric flow rate to a mass flow rate using the following equation:
Qm = Qv ρ Where:
Qm is the mass flow rate
Qv is the volumetric flow rate, and
ρis the fluid density
NOTE
A density value is required to configure the flow units for mass flow rate measurement.
Sensor Range: High
This parameter is the maximum value that the PV Range value can be set to. This is the upper measuring limit of the transmitter and sensor.
Sensor Range: Low
This parameter is the minimum value that the PV Range value can be set to. This is the lower measuring limit of the transmitter and sensor.
Measurement Mode
Enable or disable the transmitter’s ability to read reverse flow.
Measurement Mode allows the transmitter to read negative flow. This may occur when flow in the pipe is going the negative direction, or when either electrode wires or coil wires are reversed. This also enables the totalizer to count in the reverse direction.