Consult factory for installations requiring cathodic protection or situations where there are high currents or high potential in the process.
The sensor case should always be earth grounded in accordance with national and local electrical codes. Failure to do so may impair the protection provided by the equipment. The most effective grounding method is direct connection from the sensor to earth ground with minimal impedance.
The Internal Ground Connection (Protective Ground Connection) located in side the junction box is the Internal Ground Connection screw. This screw is identified by the ground symbol:
|
| Grounding Options |
| |
Type of Pipe | No Grounding Options | Grounding Rings | Grounding Electrodes | Lining Protectors |
Conductive Unlined Pipe | See Figure | Not Required | Not Required | See Figure |
Conductive Lined Pipe | Insufficient Grounding | See Figure | See Figure | See Figure |
Insufficient Grounding | See Figure | See Figure | See Figure |