Reference Manual

00809-0100-4665, Rev AA August 2010

Rosemount 8732

Model 7400 Sensor to Rosemount 8732 Transmitter

Connect coil drive and electrode cables as shown in Figure E-4.

Figure E-4. Wiring Diagram for Brooks Sensor Model 7400 and Rosemount 8732

Table E-5. Brooks Model 7400 Sensor Wiring Connections

BROOKS MODEL 7400

ROSEMOUNT 8732

TRANSMITTER

Rosemount 8732

 

Brooks Sensors Model 7400

1

 

 

 

Coils +

2

 

 

 

Coils –

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

17

 

 

Shield

18

 

 

Electrode +

19

 

 

Electrode –

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This is a pulsed DC magnetic flowmeter.

Do not connect AC power to the sensor or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary.

E-7

Page 133
Image 133
Emerson manual Model 7400 Sensor to Rosemount 8732 Transmitter, Trans MIT TER