Reference Manual
Rosemount 8732
Taylor Sensor to Rosemount 8732 Transmitter
Connect coil drive and electrode cables as shown in Figure
Figure E-20. Generic Wiring Diagram for Taylor Sensors and Rosemount 8732
Table E-21. Taylor Sensor Wiring Connections
| TAYLOR | ROSEMOUNT 8732 |
| SENSORS | TRANSMITTER |
| E2 |
|
Electrodes | E1 |
|
| S1 and S2 |
|
| Green |
|
Coils | White |
|
|
| |
| Black |
|
|
| Fuse |
Rosemount 8732 |
| Taylor Sensors | ||||
1 |
|
|
| Black | ||
2 |
|
|
| White | ||
|
|
|
|
|
| Green |
|
|
|
|
| ||
17 |
|
| S1 and S2 | |||
18 |
|
| E1 | |||
19 |
|
| E2 | |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
This is a pulsed DCDC magnetic flowmeter. Do not connect AC power to the sensor or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary.