
Rosemount 8732
Reference Manual
PROCEDURES | If the output of your Rosemount 8732 is unstable, first check the wiring and |
| grounding associated with the magnetic flowmeter system. Ensure that the |
| following conditions are met: |
| • Ground straps are attached to the adjacent flange or ground ring? |
| • Grounding rings, lining protectors, or grounding electrodes are being |
| used in lined or nonconductive piping? |
| • Both of the shields are attached at both ends? |
| The causes of unstable transmitter output can usually be traced to extraneous |
| voltages on the measuring electrodes. This “process noise” can arise from |
| several causes including electrochemical reactions between the fluid and the |
| electrode, chemical reactions in the process itself, free ion activity in the fluid, |
| or some other disturbance of the fluid/electrode capacitive layer. In such noisy |
| applications, an analysis of the frequency spectrum reveals process noise |
| that typically becomes significant below 15 Hz. |
| In some cases, the effects of process noise may be sharply reduced by |
| elevating the coil drive frequency above the 15 Hz region. The Rosemount |
| 8732 coil drive mode is selectable between the standard 5 Hz and the |
| |
| instructions on how to change the coil drive mode to 37 Hz. |
Auto Zero | To ensure optimum accuracy when using 37 Hz coil drive mode, there is an |
| auto zero function that must be initiated during |
| operation is also discussed in the |
| using 37 Hz coil drive mode it is important to zero the system for the specific |
| application and installation. |
| The auto zero procedure should be performed only under |
| the following conditions: |
| • With the transmitter and sensor installed in their final positions. This |
| procedure is not applicable on the bench. |
| • With the transmitter in 37 Hz coil drive mode. Never attempt this |
| procedure with the transmitter in 5 Hz coil drive mode. |
| • With the sensor full of process fluid at zero flow. |
Signal Processing
These conditions should cause an output equivalent to zero flow.
If the 37 Hz coil drive mode has been set, and the output is still unstable, the damping and signal processing function should be used. It is important to set the coil drive mode to 37 Hz first, so the loop response time is not increased.
The 8732 provides for a very easy and straightforward