Rosemount 8732
Reference Manual
Series 1800 (Version 2) Sensor to Rosemount 8732 Transmitter
Connect coil drive and electrode cables as shown in Figure
Figure E-13. Wiring Diagram for Foxboro Series 1800 (Version 2) and Rosemount 8732
Table E-14. Foxboro Generic Sensor Wiring Connections
FOXBORO SERIES |
| ROSEMOUNT |
1800 SENSOR |
| 8732 |
(VERSION 2) |
| TRANSMITTER |
|
| White |
|
| Black |
|
| Shield |
Electrode |
|
|
Connections |
|
|
GND | L2 | L1 |
| Coil Connections |
Rosemount 8732 |
| Foxboro Series 1800 Sensors | ||||
1 |
|
|
| L1 | ||
2 |
|
|
| L2 | ||
|
|
|
|
|
| Chassis Ground |
|
|
|
|
| ||
17 |
|
| Any Shield | |||
18 |
|
| Black | |||
19 |
|
| White | |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
This is a pulsed DC magnetic flowmeter.
Do not connect AC power to the sensor or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary.