Appendix A

Reference Data

FUNCTIONAL SPECIFICATIONS

Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . page A-1E-Series Advanced Diagnostics Capabilities . . . . . . . . . . page A-4 Output Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-4 Profibus PA fieldbus Digital Output Specifications . . . . . page A-4 Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . page A-6 Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-8 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-9

Sensor Compatibility

Compatible with Rosemount 8705, 8711, 8721, and 570TM sensors. Compatible with Rosemount 8707 sensor with D2 Dual calibration option. Compatible with AC and DC powered sensors of other manufacturers.

Sensor Coil Resistance

350 Ω maximum

Transmitter Coil Drive Current

500 mA

Flow Rate Range

Capable of processing signals from fluids that are traveling between 0.04 and 39 ft./s (0.01 to 12 m/s) for both forward and reverse flow in all sensor sizes. Full scale continuously adjustable between –39 and 39 ft./s (–12 to 12 m/s).

Conductivity Limits

Process liquid must have a conductivity of 5 microsiemens/cm (5 micromhos/cm) or greater. Excludes the effect of interconnecting cable length in remote mount transmitter installations.

Power Supply

90 -250 V AC, 50–60 Hz or 12-42 V DC

AC Power Supply Requirements

Units powered by 90-250 V AC have the following power requirements.