Rosemount 8732
Reference Manual
Index | Parameter Name | Description |
117 | METER_VERIF_TEST_COND_IN | Incoming Test Condition of the Flowmeter |
|
| Verification Test |
|
| Note: This parameter needs to be configured |
|
| before the Meter Verification is started via the |
|
| PERFORM_METER_VERIFY parameter. |
118 | METER_VERIF_TEST_COND_OUT | Actual Test Condition of the Flowmeter |
|
| Verification Test. |
119 | METER_VERIF_CRITERIA | Test limits that the Flowmeter Verification Test |
|
| was ran against. |
120 | METER_VERIF_RESULT | Flowmeter Verification Test summary result. |
121 | COIL_RESIST_RESULT | Coil Resistance Test Result, part of the |
|
| Flowmeter Verification Test. |
122 | COIL_INDUCT_RESULT | Coil Inductance Test Result, part of the |
|
| Flowmeter Verification Test. |
123 | ELECT_RESIST_RESULT | Electrode Resistance Test Result, part of the |
|
| Flowmeter Verification Test. |
124 | INT_SIM_RESULT | Internal Simulator Test Result, part of the |
|
| Flowmeter Verification Test. |
125 | Reserved | Reserved. |
126 | COIL_INDUCT_VALUE | Coil Inductance value. |
127 | COIL_INDUCT_DEVIATION | Coil Inductance Deviation. |
128 | COIL_INDUCT_FINGERPRINT | Coil Inductance Fingerprint. |
|
| Note: These values should be made Read |
|
| Only in the DD. Only reset by the factory |
|
| parameter NV_RESET, not by the RESTART |
|
| w/ defaults. |
129 | COIL_RESIST_VALUE | Coil Resistance value |
130 | COIL_RESIST_FINGERPRINT | Coil Resistance Fingerprint |
|
| Note: These values should be made Read |
|
| Only in the DD. Only reset by the factory |
|
| parameter NV_RESET, not by the RESTART |
|
| w/ defaults. |
131 | ELECT_RESIST_VALUE | Electrode Resistance value |
132 | ELECT_RESIST_FINGERPRINT | Electrode Resistance Fingerprint |
133 | INT_SIM_DEVIATION | The internal flow simulator reading as a |
|
| percent deviation from the reference value. |
134 | INT_SIM_REF_VALUE | Internal Flow Simulator Reference Value |
135 | INT_SIM_VALUE | The internal flow simulator reading. |
136 | METER_VERIF_EP_LIM | Flowmeter Verification Limit – Empty Pipe |
|
| condition. |
137 | METER_VERIF_FLOWING_LIM | Flowmeter Verification Limit – Flowing |
|
| condition. |
138 | METER_VERIF_NO_FLOW_LIM | Flowmeter Verification Limit – No Flow |
|
| condition. |
139 | RECALL_FINGERPRINT_VALUES | Perform a recall of the previous fingerprint |
|
| values. |
140 | PERFORM_REFINGERPRINT_FLO | Perform the |
| WTUBE | Note: The parameter |
|
| FINGERPRINT_SELECT needs to be |
|
| configured before the Refingerprint operation |
|
| is started. |
141 | PERFORM_METER_VERIFY | Perform the meter verification command. |
|
| Note: The parameter |
|
| METER_VERIF_TEST_SCOPE and |
|
| METER_VERIF_TEST_COND_IN needs to |
|
| be configured before the Refingerprint |
|
| operation is started. |