Fischer and Porter Sensor to Rosemount 8732 Transmitter
Connect coil drive and electrode cables as shown in Figure
Figure E-11. Generic Wiring Diagram for Fischer and Porter Sensors and Rosemount 8732
Table E-12. Fischer and Porter
Generic Sensor
Wiring Connections
FISCHER AND PORTER | ROSEMOUNT 8732 | |
SENSORS | TRANSMITTER | |
Electrodes |
|
|
| 2 |
|
| 1 |
|
Coils | 3 |
|
|
| |
| Chassis |
|
| M2 |
|
| M1 |
|
Fuse |
|
|
Rosemount 8732 |
| Fischer and Porter Sensors | ||||
1 |
|
|
| M1 | ||
2 |
|
|
| M2 | ||
|
|
|
|
|
| Chassis Ground |
|
|
|
|
| ||
17 |
| 3 | ||||
18 |
| 1 | ||||
19 |
| 2 | ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
This is a pulsed DC magnetic flowmeter.
Do not connect AC power to the sensor or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary.