![](/images/new-backgrounds/1329577/329577177x1.webp)
Reference Manual
Rosemount 8732
European Certifications
E1 ATEX Flameproof
Certificate No.: KEMA 07ATEX0073 X II 2G
Ex de IIC or Ex de [ia] IIC T6
Vmax = 250 V AC or 42 V DC
0575
ED ATEX Flameproof
Certificate No.: KEMA 07ATEX0073 X II 2G
Ex de IIB or Ex de [ia] IIB T6
Vmax = 250 V AC or 42 V DC 0575
ND(1) ATEX Dust Certificate No.: KEMA 06ATEX0006
II 1D
Ex tD A20 IP66 T 100 °C with I.S.: [Ex ia] IIC II (1)G
Amb. Temp. Limits:
Vmax = 250 V AC or 42 V DC IP 66
0575
(1)Max surface temperature is 40 °C above the ambient temperature conditions. Tmax = 100 °C
Special conditions for safe use (KEMA 07ATEX0073 X):
Contact Rosemount Inc. for information on the dimensions of the flameproof joints. The property class of the security screws which attach the flowtube or junction box to the transmitter is SST
Installation Instructions:
The cable and conduit entry devices and blanking elements shall be of a certified flameproof or increased safety type, suitable for the conditions of use and correctly installed. With the use of conduit, a certified stopping box shall be provided immediately to the entrance of the enclosure.
N1 ATEX Type n Certificate No: BASEEFA 07ATEX0203X
II 3(1)G
Ex nA nL [ia] IIC T4
with LOI:
0575
Special conditions for safe use (x)
The apparatus is not capable of withstanding the 500V insulation test required by Clause 6.8.1 of EN