
Rosemount 8732
Reference Manual
Table
Message | Local Display Error | Potential Cause | Corrective Action |
Message (English) | |||
“DSP Hardware not | Incompatible SW | DSP Software Revision is not equal to | Install software revision equal to Hornet's Expectation |
compatible with |
| Hornet’s Expectations. |
|
software” |
|
|
|
Table
| Local Display Error |
|
|
Message | Message (English) | Potential Cause | Corrective Action |
Grounding/Wiring | Grnd/Wire Fault | Improper installation of wiring | See “Sensor Connections” on page |
Fault |
| Coil/Electrode shield not connected | See “Sensor Connections” on page |
|
| Improper process grounding | See “Grounding” on page |
|
| Faulty ground connection | Check wiring for corrosion, moisture in the terminal block, |
|
|
| and refer to “Grounding” on page |
|
| Sensor not full | Verify sensor is full and empty pipe diagnostic is on |
High Process Noise | Hi Process Noise | Slurry flows - mining/pulp stock | Decrease the flow rate below 10 ft/s (3 m/s) |
|
|
| Complete the possible solutions listed under “Step 2: |
|
|
| Process Noise” on page |
|
| Chemical additives upstream of the | Move injection point downstream of the sensor, or move |
|
| sensor | the sensor |
|
|
| Complete the possible solutions listed under “Step 2: |
|
|
| Process Noise” on page |
|
| Electrode not compatible with the | Refer to the Rosemount Magnetic Flowmeter Material |
|
| process fluid | Selection Guide |
|
| Air in line | Move the sensor to another location in the process line to |
|
|
| ensure that it is full under all conditions |
|
| Electrode coating | Use bulletnose electrodes |
|
|
| Downsize sensor to increases flowrate above 3 ft/s (1 m/s) |
|
|
| Periodically clean sensor |
|
| Styrofoam or other insulating particles | Complete the possible solutions listed under “Step 2: |
|
|
| Process Noise” on page |
|
|
| Consult factory |
|
| Low conductivity fluids | Trim electrode and coil wires - refer to “Installation” on |
|
| (below 10 microsiemens/cm) | page |
Table
Message | Potential Cause | Corrective Action |
| Transmitter Calibration Verification | Verify pass/fail criteria |
| test failed | Rerun 8714i Meter Verification under no flow conditions |
|
| Verify calibration using 8714D Calibration Standard |
|
| Perform digital trim |
|
| Replace electronics board |
8714i Failed | Sensor Calibration test failed | Verify pass/fail criteria |
|
| Perform sensor test - see Table |
| Sensor Coil Circuit test failed | Verify pass/fail criteria |
|
| Perform sensor test - see Table |
| Sensor Electrode Circuit test failed | Verify pass/fail criteria |
|
| Perform sensor test - see Table |