Rosemount 8732

Reference Manual

00809-0100-4665, Rev AA August 2010

Table 6-1. Rosemount 8732 Basic Diagnostic Messages

Message

Local Display Error

Potential Cause

Corrective Action

Message (English)

“DSP Hardware not

Incompatible SW

DSP Software Revision is not equal to

Install software revision equal to Hornet's Expectation

compatible with

 

Hornet’s Expectations.

 

software”

 

 

 

Table 6-2. Rosemount 8732 Advanced Diagnostic Messages (Suite 1 - Option Code D01)

 

Local Display Error

 

 

Message

Message (English)

Potential Cause

Corrective Action

Grounding/Wiring

Grnd/Wire Fault

Improper installation of wiring

See “Sensor Connections” on page 2-11

Fault

 

Coil/Electrode shield not connected

See “Sensor Connections” on page 2-11

 

 

Improper process grounding

See “Grounding” on page 5-12

 

 

Faulty ground connection

Check wiring for corrosion, moisture in the terminal block,

 

 

 

and refer to “Grounding” on page 5-12

 

 

Sensor not full

Verify sensor is full and empty pipe diagnostic is on

High Process Noise

Hi Process Noise

Slurry flows - mining/pulp stock

Decrease the flow rate below 10 ft/s (3 m/s)

 

 

 

Complete the possible solutions listed under “Step 2:

 

 

 

Process Noise” on page 6-7

 

 

Chemical additives upstream of the

Move injection point downstream of the sensor, or move

 

 

sensor

the sensor

 

 

 

Complete the possible solutions listed under “Step 2:

 

 

 

Process Noise” on page 6-7

 

 

Electrode not compatible with the

Refer to the Rosemount Magnetic Flowmeter Material

 

 

process fluid

Selection Guide (00816-0100-3033)

 

 

Air in line

Move the sensor to another location in the process line to

 

 

 

ensure that it is full under all conditions

 

 

Electrode coating

Use bulletnose electrodes

 

 

 

Downsize sensor to increases flowrate above 3 ft/s (1 m/s)

 

 

 

Periodically clean sensor

 

 

Styrofoam or other insulating particles

Complete the possible solutions listed under “Step 2:

 

 

 

Process Noise” on page 6-7

 

 

 

Consult factory

 

 

Low conductivity fluids

Trim electrode and coil wires - refer to “Installation” on

 

 

(below 10 microsiemens/cm)

page 2-1

Table 6-3. Rosemount 8732 Advanced Diagnostic Messages (Suite 2 - Option Code D02)

Message

Potential Cause

Corrective Action

 

Transmitter Calibration Verification

Verify pass/fail criteria

 

test failed

Rerun 8714i Meter Verification under no flow conditions

 

 

Verify calibration using 8714D Calibration Standard

 

 

Perform digital trim

 

 

Replace electronics board

8714i Failed

Sensor Calibration test failed

Verify pass/fail criteria

 

 

Perform sensor test - see Table 6-5 on page 6-8

 

Sensor Coil Circuit test failed

Verify pass/fail criteria

 

 

Perform sensor test - see Table 6-5 on page 6-8

 

Sensor Electrode Circuit test failed

Verify pass/fail criteria

 

 

Perform sensor test - see Table 6-5 on page 6-8

6-4

Page 68
Image 68
Emerson Rosemount 8732 manual Message Potential Cause Corrective Action