Rosemount 8732
Reference Manual
00809-0100-4665, Rev AA August 2010
Table 6-5. Sensor Test
Test | Sensor | Required | Measuring at | Expected Value | Potential Cause | Corrective Action | |||||||||||||
Location | Equipment | Connections | |||||||||||||||||
A. Sensor | Installed or | Multimeter | 1 and 2 = R | 2Ω ≤ R ≤ 18Ω | • Open or | • Remove and | |||||||||||||
Coil | Uninstalled |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Shorted Coil |
| replace sensor |
B. Shields to | Installed or | Multimeter | 17 and |
|
|
|
| < 0.2Ω | • | Moisture in | • | Clean terminal | |||||||
|
|
|
| ||||||||||||||||
Case | Uninstalled |
|
|
|
| and case |
|
|
|
| terminal block |
| block | ||||||
|
|
|
|
|
|
|
|
| • Leaky electrode | • Remove sensor | |||||||||
|
|
|
|
|
|
|
|
| |||||||||||
|
|
| ground |
|
|
| • Process behind |
|
| ||||||||||
|
|
| 17 and case |
|
|
|
| liner |
|
| |||||||||
|
|
|
|
|
|
|
|
|
| ||||||||||
|
|
| ground |
|
|
|
|
|
|
| |||||||||
C. Coil Shield | Installed or | Multimeter | 1 and |
|
|
|
| ∞Ω (< 1nS) | • | Process behind | • | Remove sensor | |||||||
|
|
|
|
|
| ||||||||||||||
to Coil | Uninstalled |
| 2 and |
|
|
|
|
|
| ∞Ω (< 1nS) |
| liner |
| and dry | |||||
|
|
|
|
|
|
|
|
|
| • | Leaky electrode | • | Clean terminal | ||||||
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| • | Moisture in |
| block |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| terminal block | • | Confirm with |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| sensor coil test |
D. Electrode | Installed | LCR (Set to | 18 and 17 = R1 | R1 and R2 should be stable | • Unstable R1 or | • Remove coating | |||||||||||||
Shield to |
| Resistance | 19 and 17 = R2 | NO: | ≤ 300Ω |
| R values |
| from sensor wall | ||||||||||
Electrode |
| and 120 Hz) |
|
|
|
|
|
|
|
|
|
| N5, E5, CD, |
| 2 | • | Use bulletnose | ||
|
|
|
|
|
|
|
|
|
|
|
| confirm coated | |||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ED: | ≤ 1500Ω |
| electrode |
| electrodes | |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| • | Shorted | • | Repeat |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| electrode |
| measurement |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| • | Electrode not in | • | Pull sensor, |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| contact with |
| complete test in |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| process |
| Table |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| • Empty Pipe |
| Table | |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| • | Low conductivity |
| page |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| • | Leaky electrode |
| line. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
To test the sensor, a multimeter capable of measuring conductance in nanosiemens is preferred. Nanosiemens is the reciprocal of resistance.
1nanosiemens | = | 1 |
|
| 1gigaohm |
| or |
|
1nanosiemens | = | 1 |
|
| 1 × 109ohm |