DSP-100/2000

Users Manual

Selected Autotest Reports: Displays the SELECT AUTOTEST REPORTS screen where you can select the reports you want to print, as follows:

1.Use #Page Up, $Page Down, and/or DUto highlight a report label.

2.Press Eto select the highlighted report. An asterisk appears on the

label to show that the report is selected for printing. To select several consecutive reports, hold down E. To remove an asterisk, highlight the report label; then press E.

3.To print the selected reports, press ! Start Print. To stop printing and return to the main Print menu, press e.

All Report Summary: Same as All Autotest Reports, except that reports are printed in summary form. The summaries give the date and time each report was saved, the cable identification entered for each report, and an overall test result for each report.

Selected Report Summary: Same as Selected Autotest Reports, except that reports are printed in summary form.

Edit Report Identification: Displays the REPORT IDENTIFICATION screen, where you can edit the Custom Header, Operator Name, or Site Name, as follows:

1.Turn the rotary switch to PRINT.

2.Use Dto highlight Edit Report Identification; then press E.

3.Use DUto highlight the information you want to edit; then press

E.

If you are editing an operator or site name, you can press @New to add a new name. The New softkey appears only if less than 20 names have been entered.

To rename or delete an existing operator or site name, press !Edit, select the desired name; then press !Rename or @Delete. Changes to names preceeded by a "$" appear on printed test reports. You cannot delete a name that is used on a saved report.

5-4

Page 104
Image 104
Fluke DSP-2000 user manual DSP-100/2000