Autotest on Twisted Pair Cable

1.If you are using a DSP-100 main unit as the remote, turn the remote unit’s rotary switch to SMART REMOTE. If you are using a smart remote unit, turn its rotary switch to ON.

2.Use a 2m patch cable of the correct impedance to connect the remote to the far end of the cable link.

3.On Model DSP-100, remove any cable connected to the test tool’s BNC connector.

4.Turn the rotary switch on the main unit to AUTOTEST.

5.Verify that the settings displayed are correct. You can change these settings in the SETUP mode.

6.Use a 2m patch cable of the correct impedance to connect the test tool to the near end of the cable link. On Model DSP-2000, connect to the CABLE TEST jack.

7.Press T to start the Autotest.

2-9