DSP-100/2000

Users Manual

Note

To extend battery life, use the ac adapter/charger when monitoring impulse noise for extended periods. While monitoring impulse noise, the test tool’s automatic power-down function is disabled.

Note

Monitoring impulse noise without a remote connected may produce unreliable test results.

1.If you are using another main unit as the remote, turn the remote unit’s rotary switch to SMART REMOTE. If you are using a smart remote unit, turn its rotary switch to ON.

2.Use a 2m patch cable of the correct impedance to connect the remote to the far end of the cable link.

3.On Model DSP-100, remove any cable connected to the test tool’s BNC connector.

4.Turn the rotary switch to MONITOR.

5.Use Dto highlight Impulse Noise on RJ45.

6.Use a 2m patch cable of the correct impedance to connect the test tool to the near end of the cable link. On Model DSP-2000, connect to the CABLE TEST jack.

7.Press Tto start the noise test.

4-22

Page 98
Image 98
Fluke DSP-2000, DSP-100 user manual Use Dto highlight Impulse Noise on RJ45